410 Hg Analyzer

»Home »Product Andalan »BUCK SCIENTIFIC »ATOMIC ABSORPTION SPECTROPHOTOMETERS » 410 Hg Analyzer

410 Hg Analyzer

410 Hg Analyzer

The Model 410 Mercury Analyzer incorporates the Hatch & Ott Cold Vapor Technique for the analysis of trace levels of mercury in drinking, surface and saline waters, as well as domestic and industrial wastes.  Perfect for EPA Method 245.1.

Features

Detection limit 10 PPT
400 mm Flowcell for lower detection levels
Complete analytical system includes Hg Cold
Vapor Kit
Affordable price perfect for labs with limited budget
Quick analysis and recovery time
Up to eight point calibration curve
Push button Quick-Purge for flow cell flushing
RS 232, Printer and analog recorder output
Simple to use-minimal operator training
D2 Background correction
Live technical assistance for the life of the instrument

Summary of method

This method is based on the absorption of radiation by mercury vapor at the 253.7nm line. The mercury is first reduced to the elemental state using the procedure outlined in EPA method 245.1. Air is then bubbled through the sample, speeding the mercury vapor to the absorption cell. The 253.7nm mercury line emitted by the lamp is absorbed by the vapor in the cell in proportion to the mercury concentration. The result is transmitted to the Digital readout which can be read directly in concentration of mercury. The instrument has a built-in “Peak-Picker” which reads peak area. The operator records this reading before purging. The instrument is then ready for the next sample.

Specifications
Detection Limits: 10 PPT (In a clean lab environment)
Absorption Cell: 400 mm pathlength with silica windows
Power Requirements: 115/220V.A.C, 50/60Hz, 80 watts
Dimensions: 39"W x 11"D x 12"H  Weight: 45 lbs.
Output:: RS232 and Analog
Background Correction: In-line Deuterium Lamp
Delivery system: Includes air dryer and charcoal filter