NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2

»Home »Product Andalan »GENTECH »Natural Gas Analysis » NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2

NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2

NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2

The NGA-008 Natural Gas Analyzer for Hydrocarbons, CO, and CO2 is configured with two channels - Flame Ionization Detector (FID), packed and caplillary columns, three valves, methanizer, and an auxiliary oven. The NGA-008 meets the following official methods.

  • ASTM D2505 - 88 - Standard Test Method for Ethylene, Other Hydrocarbons, and Carbon Dioxide in High-Purity Ethylene by Gas Chromatography
  • UOP603 - 13 - Trace CO and CO2 in Hydrogen and Light Gaseous Hydrocarbons by GC

GenTech Master Gas Chromatograph (GC) offers a new approach to the world of gas chromatography by dramatically decreasing sample run times in a wide range of GC Applications.

This GC is truly customizable to meet each user’s needs, while maintaining ease of use.  Its modular design allows the user to easily change or upgrade to any GC configuration.  Up to three injection units and three detectors can be mounted simultaneously.  An intuitive Touch Screen interface controls both the GC and AS with ultimate simplicity. The whole system is completely accessible - allowing easy maintenance and reducing cost.

Features:

NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2GenTech Master Gas Chromatograph (GC)

  • User-Friendly Touch Screen Interface
  • DPC – Patented Digital Flow Control
  • Constant Linear Velocity Mode
  • Pulsed Injection
  • Gas Saving and Leak Checking System
  • Reduced Sample Run Times
  • Complete Range of Highly Sensitive Detectors
  • Three Injection Ports and Three Detectors
  • Clarity™ Work Station - compatibility
  • Versatile System Design
  • Column Oven 1 - 500° C
  • Analytical Methods Capacity of 25
  • Digital Pneumatic Control Pressure Range of up to 120 psi
  • Typical Cooling Time of 4 Minutes
  • Narrow Bore of Fast GC Columns of 50 - 2000 mm I.D.
  • Typical Peak Widths of 0.5 Seconds or Less
  • Conventional GC Application is 161.228 (split 2,20)
  • Fast GC Application
  • Signal-to-Noise Ratio of 2,2´,5´
  • Increased Sensitivity

NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Clarity™ Software Control

  • Clarity™ is an advanced chromatography station designed to control up to four instruments and to acquire and evaluate up to four analog signals from each instrument simultaneously
  • Includes an A/D converter
  • Clarity™ features a user interface that is simple and intuitive
  • Guarantees security and control of chromatographic data according to GLP and FDA 21 CFR Part 11 standards.

Includes:

NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2MANUFACTURED IN 2013!
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2ONE (1) YEAR WARRANTY INCLUDING APPLICATION SUPPORT
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Complete Data System Control using Clarity software. Computer with Software License
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2GenTech Master GC
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Dual Flame Ionization Detectors (FID) with Digital Flow Control
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Split/Splitless Injector
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Three Valves - Two (2) 6 port and One (1) 8 port, air actuated gas sampling valves
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Dual total flow modules for capillary inlet
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Methanizer
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Auxiliary Oven
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO21 ml, 250 μl and 500 μl sample loops
NGA-008 Gas Analyzer for Hydrocarbons, CO, and CO2Two (2) 0.32 mm capillary columns and Two (2) 1/8" packed columns for NGA-008 Analysis

GenTech Master Gas Chromatograph (GC) Specifications

Column Oven:

Dimensions Width Height Depth

  280 mm 280 mm 160 mm

Operating Temperature

5°C above ambient to 500°C
-50°C to 500°C with liquid CO2 cryogenic cooling option
-100°C to 500°C with liquid N2 cryogenic cooling option

Temperature Resolution

0.1°C

Thermal Accuracy

± 0.1°C

Thermal Stability

± 0.1°C

Temperature Programming

25 ramps, 26 isotherms

Max Heating Ramp Rate

140°C/min, 0.1 °C/min resolution

Typical Cool-Down Rate

300°C to 50°C in 4 min

Isothermal Time

0.00 to 999.00 min, 0.01 min resolution

Pneumatic Controls:

Pressure Range

0-120 psi

Injector Carrier Gas Max Total Flow

1000 mL/min (He)

Injector Pressure/ Flow Programming

Constant or programmed flow
Constant or programmed pressure
Constant linear velocity
Pulsed Injection
Flow programming: 25 ramps/26 cost. flows
Pressure programming: 25 ramps/26 isobars
Pressure resolution: 0.01 bar

Atmospheric Pressure and Temperature Compensation

Yes

Injectors Carrier Gas Control

Digital Flow Control (DFC)

Injectors Split Gas Control

Digital Flow Control (DFC)

Injectors Purge Gas Control

Digital Flow Control (DFC)

Injectors Gas Saver

Yes

Detectors Gas Control

Digital Pressure Control

Injectors:

Injectors Available

Packed (PK)
Split/splitless (SL/IN)
Programmable Temperature Vaporizer (PTV)
Gas sampling valves (GSV)

Carrier Gas

He, N2, H2, Ar, Ar+CH4

Maximum Number Installed Simultaneously

3

Maximum Number of PTV

2

Detectors:

Detectors Available

FID     flame ionization detector
NPD    nitrogen and phosphorous detector
ECD    electron capture detector
PID     photo ionization detector
TCD    thermal conductivity detector
µTCD  micro thermal conductivity detector
FPD     flame photometric detector

Maximum Number Installed Simultaneously

3

Maximum Number of TCD

2

Maximum Number of µTCD

2

Output Signals

Digital (LAN, USB, RS-232) and Analog outputs (0-1V, 0-10V) 

Data Communication:

Remote Control

Start in; start and ready out (contact relays normally open)

Communication

RS232, LAN, USB

Time Events

n.4 voltage free contacts (contact relays normally open) n.4 24V outputs

Touch Screen Graphic User Interface

Display

LCD TFT, 5.7”

Resolution

240 x 320 pixel

Colors

65,536

Various:

Inside Oven Lamp

Yes

Analytical Methods Capacity

25